Vacuquick

Vacuquick 13-B

No tap-water connection necessary,
Chamber: ř 18 x 23 cm, volume: 5.7 litres

Vacuquick 14-B

Tap-water connection necessary,
Chamber: ř 18 x 23 cm, volume: 5.7 litres


Previously, the sterilization of handpieces between two patients was only required for invasive work. Several hygienic experts recommend the hygienic preparation of handpieces after every patient. And in fact also in principle a disinfection of the inner parts is required; sterilization is necessary for invasively used handpieces.

For this demand an autoclave with very short operation time is required. The quick program for unwrapped material to be sterilized takes only 7 minutes for the Vacuquick! And that is achieved even though a fractionated vacuum is used for sterilization! The Vacuquick autoclaves combine in one device the advantages of a quick autoclave with those of a compact practice autoclave with very modest space requirements for the installation.

The Vacuquick autoclaves are large enough that when loading 4 trays in continuous operation in an eight hour day in the practice, more than 70 kg of wrapped instruments can be sterilized. The Vacuquick autoclaves are operated "intuitively": All information about the autoclave as well as the program sequence is displayed on the graphic display (8 cm width x 6 cm height). This avoids operating errors.

Vacuquick 13-B is a "stand alone" autoclave with integrated water storage tank and patented air cooling system, and is ready for operation without further Installation work. The Vacuquick 13-B is delivered with an external condensate container for the used water.

Vacuquick 14-B requires a tap-water connection with the advantage that if it is connected to the water treatment system MELAdem 40, it is automatically supplied with high quality demineralized water. If many sterilization cycles are carried out daily, the MELAdem 47 is recommended. The tap-water connection in conjunction with the high-performance vacuum pump has only a very low heat emission.

Vacuquick 13-B and Vacuquick 14-B are "Class B" autoclaves and thereby comply with the highest requirements of the European Standard EN 13060.
Whether sterilizing wrapped instruments or unwrapped material, sterilization is always conducted with the greatest safety. Thanks to its compact exterior dimensions, the Vacuquick autoclaves of the Premium-Class also fit into the openings of older built-in cupboards, so that a "Class B" autoclave can also be installed in these practices in the same familiar spot.

The required documentation of the sterilization cycles is carried out over the built-in network interface. The data can be stored directly on the practice network. It is also possible to print the data with the protocol printer MELAprint 42 or store the data on the MELAflash CF Card. The sterilization records can be read out on every PC with the MELAflash CF Card reader.

aTwo helix reference block systems, the MELAcontrol and MELAcontrol PRO are available for the verification of successful sterilization in a class B autoclave.

aThe documentation of the release is carried out with the MELAdoc system, consisting of the documentation pages and the MELAdoc label printer for sterilization packagings.

For an optimum loading of the chamber, the Vacuquick is delivered with a tray mount for 4 trays or for 2 half-tray-cassettes. Likewise different sterilization containers are available. 2 pieces of the container 17K fit into the chamber of the Vacuquick.

Technical Data:

 

  Vacuquick 13-B stand alone Vacuquick 14-B Tap-water connection necessary
Outside dimensions: Width: 35 cm x Height: 40 cm x Depth: 62 cm Width: 35 cm x Height: 40 cm x Depth: 57 cm
Chamber: ř 18 x 23 cm, Volume: 5.7 litres ř 18 x 23 cm, Volume: 5.7 litres
Maximum load: 4 kg instruments or 0.7 kg textiles 4 kg instruments or 0.7 kg textiles
Weight: 39 kg 41 kg


Power supply:
230 Volt / 50 Hz / 2,900 Watt

 

Program Cycle times


Vacuquick 13-B Warm start 0,5 kg Instruments Warm start 4,0 kg Instruments Warm start 0,7 kg
Textiles
Universal-Program
wrapped instruments (134°C / 2 bar)
13 minutes 19 minutes 17 minutes
Quick-Program„B“
singlewrapped instruments
(max. 1,5 kg) (134°C / 2 bar)
11 minutes 16 minutes 15 minutes
Quick-Program „S“
unwrapped instruments (134°C / 2 bar)
7 minutes 10 minutes -
Gentle-Programm
wrapped instruments (121°C / 1 bar)
25 minutes 33 minutes 29 minutes
Prion-Program
wrapped instruments (134°C / 2 bar)
30 minutes 35 minutes 33 minutes
Bowie&Dick-test
(loaded with test-indicator) (134°C / 2 bar)
approx. 15 minutes - -
Vacuum-test
(without load)
approx. 16 minutes - -


Vacuquick 14-B Warm start 0,5 kg Instruments Warm start 4,0 kg Instruments Warm start 0,7 kg
Textiles
Universal-Program
wrapped instruments (134°C / 2 bar)
12 minutes 15 minutes 14 minutes
Quick-Program„B“
singlewrapped instruments
(max. 1,5 kg) (134°C / 2 bar)
10 minutes 14 minutes 13 minutes
Quick-Program „S“
unwrapped instruments (134°C / 2 bar)
7 minutes 9 minutes -
Gentle-Programm
wrapped instruments (121°C / 1 bar)
24 minutes 28 minutes 26 minutes
Prion-Program
wrapped instruments (134°C / 2 bar)
29 minutes 33 minutes 31 minutes
Bowie&Dick-test
(loaded with test-indicator) (134°C / 2 bar)
approx. 15 minutes - -
Vacuum-test
(without load)
approx. 16 minutes - -


All cycle-times plus drying: Universal-Program / Gentle-Program / Prion-Program 15 minutes,
Quick-Program„B“ 6 minutes, Quick-Program„S“ 1 minute

Download

Vacuquick Brochure PDF .